Model 200/1 Hydrocarbon
Analyzer -
FID Methane / Non Methane (NMHC), Oven Heated Hydrocarbon
Analyzer

The VIG Industries, Inc. Model 200/1 is a microprocessor
based, oven heated methane /
non methane (NMHC) hydrocarbon analyzer for high accuracy,
sensitivity and stability. The Model 200/1 measures methane and
non methane hydrocarbons at preset intervals. The VIG Model 200/1 uses one(1)
Flame Ionization Detector (FID) to obtain methane and non-methane hydrocarbon readings. A sample is fed to the analyzer via
an
internal heated pump. After the sample reaches the analyzer, a
portion of the sample is trapped and fed through a GC (gas chromatography)
column to separate
the methane from the total and is fed to the FID detector for analysis. As soon
as the methane is read, the remaining sample is back flushed
through the GC column and fed to the FID detector for analysis of the
non-methane component. The LCD display indicates the amount of methane and non-methane hydrocarbons.
The Model 200/1 utilizes one flame
ionization detector (FID). The detector, all sample lines, GC column and all solenoid
valves that come in contact with the sample are maintained in a
controlled, heated oven to prevent condensation, and to provide
reliable performance in the analysis of concentrations of wide
varieties of hydrocarbons in gaseous mixtures and in ambient air.
Download 1.3MB
Applications
- Compliance Monitoring - US EPA
Method 18
-
Process Monitoring - Continuous
monitoring and alarm or control of:
-
Process gas streams utilizing organic
solvents
-
Crude oil
-
Other chemicals containing hydrocarbons
-
Efficiency Monitoring - Monitoring
effluent of volatile organic compound (VOC) reduction equipment for:
-
Environmental compliance
-
Efficiency control of incinerators (Thermal
or catalytic)
-
Scrubbers
-
Carbon absorbers
-
Monitoring of catalytic converters
-
Other abatement equipment
-
Safety Monitoring - Lower explosive
limit (LEL) monitoring and/or control of:
-
Ovens/dryers
-
Fugitive emissions monitoring
-
Personnel work area monitoring
-
Leak detection of process equipment or
solvent storage areas
-
Stack Monitoring
Features
- Easy to use software
- GC technology used for separation
- Automatic startup / ignition
- Automatic fuel shut-off system
- Automatic flame-out indicator
- Adjustable alarm and oven settings
- Precision 1% of full scale
- Fast response
- Methane and Non methane - minimum 2.5 minute cycle
- Two stage sample filter
with exchangeable sintered stainless steel elements
- Teflon® isolated
detector (FID)
- 19" rack or bench mount
- Heated sample pump
head
- Outputs 0-1, 0-5, or 0-10Volts
- Automatic ignition
Options
- Zero and calibration input ports with software
- 4-20mA output - no extra charge
- Range recognition relays
- RS-232 interface
- Internal combustion air supply
- Automatic sample shut-off system
Related
Available Equipment
- External zero air generator (reduces bottles)
- Hydrogen Generator (reduces bottles)
- Heated sample lines and controllers
- Strip chart recorders and data loggers
- NEMA rated enclosures
Standard Specifications
| |
|
Analysis Method
|
Oven Heated, Flame Ionization Detector (FID)
|
|
Separation Method
|
Column
|
|
Precision
|
+/- 1% of Full Scale
|
|
Response Time
|
- Methane - Approximately 40 seconds,
updated every 3 minutes
- Non methane - Approximately 70 seconds,
updated every 3 minutes
|
|
Zero Drift
|
+/- 1% of Full Scale Per 24 Hours
|
|
Span Drift
|
+/- 1% of Full Scale Per 24 Hours
|
|
Linearity
|
Within 1% of full scale
|
|
Oxygen Synergism
|
Within 1% of full scale within selected range
|
|
Ranges
|
(4 Ranges per amplifier, two amplifiers per analyzer,
one amplifier for total hydrocarbons,
one amplifier for methane and non-methane)
Standard Ranges
- 0-100, 1000, 10000, 100000ppm (LDL 0.1ppm)
- 0-10, 100, 1000, 10000ppm (LDL 0.01ppm)
|
|
Display
|
Graphic,
backlit, 240W x 64H pixels, high contrast,
wide viewing angle
|
|
Outputs
|
0-1, 5, or 10 Volts & optional 4-20mA
|
|
Alarms
|
Flame Out & Concentration, Latching
|
|
Safety
|
Flame Out Indicator with Fuel Shut-Off System
|
|
Oven Temperature
|
Operating Temperature 275°F, Adjustable 200 to 300° F
|
|
Sample Pump
|
Heated 4 Liters/Minute standard or Heated 14 Liters/Minute optional
|
|
Ignition |
Automatic (can be set to manual by operator) |
|
Glow Plug |
Main and spare glow plug, selectable by switch on back panel |
|
Dimensions |
19" front panel, 16.75" wide chassis, 18" deep chassis,
21" deep with fittings and handles, 9" high |
|
Weight |
55 to 65 pounds depending on options |
|
Operating Requirements
| |
|
Fuel
|
Hydrogen (UHP Grade) @ 18psi incoming pressure
|
|
Combustion Air |
Zero Grade Air @ 18PSI |
|
Operation Air |
Compressed Air (Instrument Air - Dry / Oil-Less) @
60PSI |
|
Zero Gas
|
Zero Air (UHP Grade) @ 10PSI
|
|
Calibration Gas |
User Selected @ 10PSI |
|
Carrier Gas
|
Nitrogen (UHP Grade) @ 30PSI
|
|
Power |
115Vac, 60Hz or optional 220Vac, 50Hz |
|
Ambient Temperature |
32 to 110º F |
|
Warranty
All instruments sold by VIG Industries, Inc., are warranted for a
period of one (1) year from the date of purchase, against defects in
materials and workmanship. VIG warrants that the product supplied
conforms to the specifications assigned thereto. There is no other
warranty either implied or expressed. Seller liability is limited
specifically to the cost or assigned value of the items sold.
|
 |