Model 210 Hydrocarbon Analyzer -
FID
Oven Heated Methane / Ethane / Non Methane Ethane (NMEHC)
/ Total Hydrocarbon Analyzer

The VIG Industries, Inc. Model 210 is a microprocessor
based, oven heated methane / ethane
/ non methane-ethane (NMEHC) total hydrocarbon analyzer for high accuracy,
sensitivity and stability. The Model 210 measures methane, ethane, non methane-ethane hydrocarbons at preset
intervals and continuously measures total hydrocarbons in real time. The Model
210 uses two (2)
Flame Ionization Detectors (FID). One FID is used to obtain methane,
ethane and non methane ethane hydrocarbon readings. The second FID is used to
continuously measure total hydrocarbons. A sample is fed to the analyzer via
an
internal heated pump. After the sample reaches the analyzer it is fed to the
first FID to measure total hydrocarbons. A small portion of the sample is trapped and fed through a
GC (gas chromatography) column to separate
the methane and ethane from the total and is fed to the second FID detector for analysis. As soon
as the methane and ethane components are read, the remaining sample is back flushed
through the column and fed to the FID detector for analysis of the
non methane ethane component. The LCD display indicates the amount of methane,
ethane, residual (non methane ethane) and total hydrocarbons.
The Model 210 utilizes two flame
ionization detectors (FID). The detectors, all sample lines, GC column and all solenoid
valves that come in contact with the sample are maintained in a
controlled, heated oven to prevent condensation, and to provide
reliable performance in the analysis of concentrations of wide
varieties of hydrocarbons in gaseous mixtures and in ambient air.
Download 1MB
Applications
- Compliance Monitoring -
U.S. E.P.A. Method 18 and Method 25A
-
Process Monitoring - Continuous
monitoring and alarm or control of:
-
Process gas streams utilizing organic
solvents
-
Crude oil
-
Other chemicals containing hydrocarbons
-
Efficiency Monitoring - Monitoring
effluent of volatile organic compound (VOC) reduction equipment for:
-
Environmental compliance
-
Efficiency control of incinerators (Thermal
or catalytic)
-
Scrubbers
-
Carbon absorbers
-
Monitoring of catalytic converters
-
Other abatement equipment
-
Safety Monitoring - Lower explosive
limit (LEL) monitoring and/or control of:
-
Ovens/dryers
-
Fugitive emissions monitoring
-
Personnel work area monitoring
-
Leak detection of process equipment or
solvent storage areas
-
Stack Monitoring
Features
- Easy to use software
- GC technology used for separation
- Automatic ignition
- Automatic fuel shut-off system
- Automatic flame-out indicator
- Adjustable alarm and oven settings
- Precision 1% of full scale
- Fast response
- Methane, Ethane and Non methane -
minimum 3.0 minute cycle
- Total Hydrocarbons - Within 5 seconds
to 90% of final reading
(continuous real time reading)
- Two stage sample filter
with exchangeable stainless steel elements
- Teflon® isolated
detector
- 19" rack or bench mount
- Heated Sample pump
- Outputs 0-1, 0-5, or 0-10Volts
Options
- Zero and calibration input ports with software
- 4-20mA output - no extra charge
- Range recognition relays
- RS-232 interface
- Internal combustion air supply
- Automatic sample shut-off system
Related
Available Equipment
- External zero air generator (reduces bottles)
- Hydrogen Generator (reduces bottles)
- Heated sample lines and controllers
- Strip chart recorders and data loggers
- NEMA rated enclosures
Standard Specifications
| |
|
Analysis Method
|
Flame Ionization Detector (FID)
|
|
Separation Method
|
Column
|
|
Precision
|
+/- 1% of Full Scale
|
|
Response Time
|
- Total Hydrocarbons - Within 5 seconds to 90% of final reading
(continuous real time reading)
- Methane - Approximately 40 seconds,
updated every 3 minutes
- Ethane - Approximately 55 seconds,
updated every 3 minutes
- Non methane - Approximately 90 seconds,
updated every 3 minutes
|
|
Zero Drift
|
+/- 1% of Full Scale Per 24 Hours
|
|
Span Drift
|
+/- 1% of Full Scale Per 24 Hours
|
|
Linearity
|
Within 1%
|
|
Oxygen Synergism
|
Within 1% of Selected Range
|
|
Ranges
|
Standard Ranges (4 Ranges per amplifier, two amplifiers per analyzer,
one amplifier for total hydrocarbons,
one amplifier for methane, ethane
and non methane-ethane)
- 0-100, 1000, 10000, 100000ppm (LDL 0.1ppm)
- 0-10, 100, 1000, 10000ppm (LDL 0.01ppm)
|
|
Display
|
Backlit Digital / Graphical LCD
|
|
Outputs
|
0-1, 5, or 10 Volts and 4-20mA.
|
|
Alarms
|
Flame Out & Concentration, Latching
|
|
Safety
|
Flame Out Indicator with Fuel Shut-Off System
|
|
Oven Temperature
|
Operating Temperature 275°F, Adjustable 200 to 300°F
|
|
Sample Pump
|
Heated 4 Liters/Minute standard or Heated 14 Liters/Minute optional
|
|
Ignition
|
Automatic (can be set to manual by operator)
|
|
Glow Plug
|
Main and spare glow plug, selectable by switch on back panel
|
|
Dimensions
|
19" front panel, 16.75" wide chassis, 18" deep chassis,
21" deep with fittings and handles, 9" high
|
|
Weight
|
55 to 65 pounds depending on options
|
|
Operating Requirements
| |
|
Fuel |
Hydrogen (UHP Grade) @ 18psi incoming pressure |
|
Combustion Air |
Zero Grade Air @ 18PSI |
|
Operation Air |
Compressed Air (Instrument Air - Dry / Oil-Less) @
60PSI |
|
Zero Gas |
Zero Air (UHP Grade) @ 10PSI |
|
Calibration Gas |
User Selected @ 10PSI |
|
Carrier Gas |
Nitrogen (UHP Grade) @ 30PSI |
|
Power |
115Vac, 60Hz or optional 220Vac, 50Hz |
|
Ambient Temperature |
32 to 110º F |
|
Warranty
All instruments sold by VIG Industries, Inc., are warranted for a
period of one (1) year from the date of purchase, against defects in
materials and workmanship. VIG warrants that the product supplied
conforms to the specifications assigned thereto. There is no other
warranty either implied or expressed. Seller liability is limited
specifically to the cost or assigned value of the items sold.
|
 |